High Voltage Power Supplies
Lynx DSA User's Manual - 9240227J
147
High Voltage Power Supplies
All outputs are current limited and short-circuit protected.
HV INHIBIT INPUT – has two modes.
•
POSITIVE POLARITY – For CANBERRA preamplifiers. Enable
condition (cold detector) is an open circuit or active high = +1.2 V to +24
V; Inhibit condition (warm detector) is –24 V to <+1.2 V or ground.
•
NEGATIVE POLARITY – For preamplifiers and LN monitors, where
Enable condition (cold detector) is –24 V to <+1.2 V; Inhibit condition
(warm detector) is open circuit or active high = +1.2 V to +24 V.
With Negative polarity selected, an open input will disable the high voltage.
Individual connectors for positive (HV+) and negative (HV–) high range (1500 V)
outputs. A separate connector is provided for the ±200 V output.
HVPS Range 1
The following applies to HVPS 1:
•
OUTPUT –
±
200 – 1500 V at 1 mA max.
•
RIPPLE – 5 mV P-P.
•
TEMPERATURE COEFFICIENT –
±
50 ppm/
°
C.
•
STABILITY – 0.01%/h, 0.02%/8 h.
•
ACCURACY – ±5%. ±2.5% Typical.
•
LOAD REGULATION – 1%.
•
SETTING RESOLUTION – 12-bit (1/4096).
HVPS Range 2
The following applies to HVPS 2:
•
OUTPUT –
±
1500–5000V at 1
µ
A max.
•
RIPPLE – 10 mV P-P.
•
TEMPERATURE COEFFICIENT –
±
50 ppm/
°
C.
•
STABILITY – 0.01%/h, 0.02%/8 h.
•
ACCURACY – ±5%. ±2.5% Typical.
•
LOAD REGULATION – 1%.
•
SETTING RESOLUTION – 12-bit (1/4096).