Chapter 5 Web-Based Operations
78
Lynx DSA User's Manual - 9240227J
Acquisition Control Panel
The Acquisition Control Panel consist of the Status & Fault indicators, Input and
Group control, Start/Stop/Clear control, Progress Bar and Acquisition Status. A brief
explanation of each follows; more details can be found in the
Lynx Web Client User's
Manual.pdf
.
Figure 42: Acquisition Panel
Status & Fault LEDs
From left to right, the LEDs indicate the state of Acquisition, High Voltage Power
Supply, Spectrum Stabilizer, and Auto Pole Zero.
•
Acquisition
indicates the state of the acquisition for the currently selected
input.
•
High Voltage Power Supply (HVPS)
shows the state of the high voltage
power supply.
•
Stabilizer
shows the state of the spectrum stabilizer.
•
Pole Zero
shows the state of automatic pole zero.
The LEDs change color based on state. The meaning of each color is as follows:
•
Red
indicates off.
•
Green
means active or waiting.
•
Yellow
means a fault has occurred with that particular subsystem.
•
Orange
means waiting for the action to happen.
•
Grey
means the feature is not supported.
During a fault condition a detailed description is displayed with a tooltip when the
mouse hovers over that LED. Holding the shift key while double clicking on the LED
will clear the fault.
Double clicking on any of the LEDs will display its settings in the Navigation Panel
to the left.
Содержание Lynx DSA
Страница 1: ...Lynx Digital Signal Analyzer 9240227J User s Manual...
Страница 10: ...viii Lynx DSA User s Manual 9240227J Notes...
Страница 232: ...222 Lynx DSA User s Manual 9240227J Notes...