Maintenance Options
Lynx DSA User's Manual - 9240227J
123
Diagnostics
This option lets you view diagnostic information.
Status
The
Status
tab displays the following read-only settings.
Figure 67: Status Diagnostics Settings
System Error
Displays the hexadecimal error code associated with the system.
PHA Input Error
Displays the hexadecimal error code associated with the PHA input.
MCS Input Error
Displays the hexadecimal error code associated with the MCS input.
System Fault
Displays any fault codes. Faults typically latch, meaning that once a fault
develops the message will be retained until acknowledged and cleared.
In most cases fault codes can be acknowledged and cleared with the
Clear
Faults & Errors
command (see below), once the condition that caused the fault
has been resolved. Persisting faults indicate that the condition causing the fault
may not have been resolved, or that service is required.
Clear Faults & Errors
Click on the
Execute
button to clear all faults and errors. Acknowledges and
clears the currently latched faults and errors. Persisting faults indicate that the
condition causing the fault may not have been resolved, or that service is
required.
Содержание Lynx DSA
Страница 1: ...Lynx Digital Signal Analyzer 9240227J User s Manual...
Страница 10: ...viii Lynx DSA User s Manual 9240227J Notes...
Страница 232: ...222 Lynx DSA User s Manual 9240227J Notes...