Lynx Genie 2000 Acquisition Window Adjust
Lynx DSA User's Manual - 9240227J
55
If acquisition has been deferred as result of the “Changer is Ready” condition not
having been met, then Genie’s acquisition status will be reported as “Wait” instead of
“Busy”.
Selection for
Ready
and
Advance
must be made according to the hardware
requirements to which the Lynx’s sample changer is being interfaced to.
Advance
Used by the
Internal Chgr
sample changer selection. Controls the DC quiescent state
of the rear panel’s CHGR ADV sample changer advance output signal as well as the
polarity of the advance pulse transition. An advance pulse will be produced when a
Changer Advance command is received through Genie’s manual or batch control.
Pulse width is always 150 milliseconds. Positive selects a TTL-low quiescent state
and produces a TTL-low-to-high pulse. Negative selects a TTL-high quiescent state
and produces a TTL-high-to-low pulse.
Selection for
Ready
and
Advance
must be made according to the hardware
requirements to which the Lynx’s sample changer is being interfaced to.
Gamma Products Sample Changer
The Gamma Products Sample Changer settings selection for the Lynx for either a
PHA or MCS definition contains the following controls.
Figure 27: Settings for Gamma Products Sample Changer
Reset/Start
This operation sends a Reset command to the changer to attempt to unjam the
changer or clear a timeout.
Load Sample/Start
The Load operation loads the next sample into the sample changer.
Sample Num.
The Load operation loads the specific sample number specified through the Sample
Num setting into the sample changer.