Chapter 5 Web-Based Operations
94
Lynx DSA User's Manual - 9240227J
The “Start Acquire” command (i.e. pressing
Start
button) arms the acquisition
electronics in the Lynx, and if the External Control is disabled (
None
is selected),
then data storage begins immediately. When the External Control is set to anything
other than
None
, then the storage of data is conditioned by the setting of the External
Start/Stop Control, as described in the next section.
None
None
disables external start/stop control. Signals at the rear panel
PHA
(or
MCS
)
ACQ SS
connector have no effect. Data storage will begin when a Start
command is received, and stop when a Stop command is received or preset
condition has been met.
Start Only
This functionality operates on the TTL-pulse edge. If the acquisition has been
armed, the edge of a TTL pulse (positive edge for
Positive
polarity, or negative
edge for
Negative
polarity) will enable storage of data. Once storage has been
enabled, subsequent pulses on the rear panel connector will have no effect. Stop
command or preset condition will terminate acquisition.
Stop Only
This functionality operates on the TTL-pulse edge. Data storage will begin as
soon as the acquisition has been armed (i.e. Start command received). Thereafter,
the edge of a TTL pulse (positive edge for
Positive
polarity, or negative edge for
Negative
polarity) will disable storage of data and terminate the acquisition.
Once storage has been stopped, subsequent pulses on the rear panel connector
will have no effect. Stop command or preset condition will also terminate
acquisition.
Start and Stop
This functionality operates on the TTL-pulse edge. If the acquisition has been
armed, the edge (positive edge for
Positive
polarity, or negative edge for
Negative
polarity) of the
first
TTL pulse will enable storage of data. Once
enabled, the edge (positive edge for
Positive
polarity, or negative edge for
Negative
polarity) of the
second
TTL pulse will disable storage of data and
terminate acquisition. Stop command or preset condition will also terminate
acquisition. Subsequent pulses on the rear panel connector once acquisition has
been terminated will have no effect.
Suspend
This functionality operates on the TTL-level. If the acquisition has been armed, a
TTL-High (for
Positive
polarity) or TTL-Low (for
Negative
polarity) will
enable
storage of data. Once enabled a TTL-Low (for
Positive
polarity) or TTL-High for
(
Negative
polarity) will
disable
storage of data. As long as the acquisition
remains armed, data storage will depend on the TTL-level as described. Stop
command or preset condition will terminate acquisition; once terminated, TTL-
level changes on the rear panel connector will have no effect.