Chapter 4 System Setup
50
Lynx DSA User's Manual - 9240227J
LT Trim
Allows adjustment of the trapezoid pulse evolution time or dead time to optimize
LTC performance. The adjustment range is 0 to 1000; the default value of 500
provides good LTC performance for a wide range of applications.
Refer to
PUR/LTC Operation
on page 189 for more information.
LTC Mode
Sets the amplifier’s Pulse Pileup Rejector and Live Time Corrector. When PUR is
On
, the pileup rejector and live time corrector (LTC) are enabled.
Off
disables the
pileup rejector and LTC. For normal operation, this setting should be
On
.
Refer to
PUR/LTC Operation
on page 189 for more information.
Gain Attn.
Sets the Gain attenuator to
Off
or
On
.
When
On
(selected) enables a divide by four input attenuator to minimize overload
due to preamp signals with large DC offsets and Reset Preamps with large output
ramp dynamic range. When
Off
is selected the signal attenuation is removed.
When making a gain adjustment, please refer to the applicable guideline under
Coarse Gain setting.
LLD
The digital Lower Level Discriminator (LLD) sets the minimum input acceptance
level; active only when Manual LLD mode is selected; adjustment range is 0.0% to
100% of the spectrum full scale range.
LLD Mode
This adjustment selects Automatic and Manual LLD modes; the digital Lower Level
Discriminator selects minimum input acceptance level. With
Auto
selected the LLD
cutoff is automatically optimized just above the spectral noise threshold.
Manual
allows the LLD cutoff to be set manually using the LLD slider bar as a percentage of
the full-scale spectral size or range. The LLD slider bar does not function when
Auto
is selected.
ULD
The digital Upper Level Discriminator (ULD) sets the maximum input acceptance
level. Range is 0 to 100%. The ULD setting is a percentage of the full-scale spectral
size or range.
TRP Gate Pol.
Selects either
Positive
or
Negative
input polarity for the TRP Gate input signal
applied to the rear panel connector.
TRP Inhibit
This setting is in effect when the TRP Inhibit Mode is set to Manual. Sets the TRP
Inhibit Time to a value from 0 to 160 μs in steps of 10 μs.