Chapter 4 System Setup
26
Lynx DSA User's Manual - 9240227J
Rack-Mount Option
The Model Lynx-Mount Kit option includes hardware for mounting a single Lynx, or
two Lynx units side by side, in 2U of a 19 inch equipment rack.
Refer to the
Rack Mounting Your Lynx Appendix
on page 202 for rack-mounting
instructions.
Genie 2000 Analysis Software
The Lynx requires Genie 2000 Basic Spectroscopy Software, Apex, or ProSpect for
acquiring and analyzing spectra data. This section concentrates on Genie 2000
operation.
Genie 2000 supports only PHA and MCS acquisition modes. Advanced acquisition
modes such as MSS, LIST, and TLIST, or the use of Auxiliary Counters and SCA
Counters are not supported by Genie.
The Lynx does not require the Genie 2000 software for acquisitions or its browser
interface.
When using Genie 2000 as the primary interface to control the Lynx, the use of the
browser interface to control the same Lynx is discouraged to prevent interference
with each other. In that case the browser interface can open the Lynx in read-only
mode for viewing purposes.
Note: Most of the information presented in this section is a review of information
presented in the Genie 2000 Operations Manual. As it may be relevant, some
areas focus on details specific to the Lynx DSA, and some details that are not
relevant to Lynx are omitted.
Installing Genie 2000 On Your PC
To install the optional Genie 2000 software on your PC, refer to the documentation
provided with the Genie 2000 software. Genie 2000 software installation is covered
in the appendix,
Software Installation
, of the
Genie 2000 Operations Manual
.
Configuring Spectrum Refresh Rate
Genie software uses the same registry parameter for configuring the default spectrum
refresh interval for the Lynx, Osprey, and DSA-LX MCAs. The default rate is 1
second. The value of this parameter must be specified in milliseconds (e.g. for a 1.5
seconds update rate enter 1500). This value should be selected according to the
capabilities of your PC; faster PC’s can tolerate smaller values for faster updates.
This property is located in Windows Registry under the following path:
Windows 32-bit:
HKEY_LOCAL_MACHINE\SOFTWARE\Canberra Industries, Inc.\Genie-2000
Environment\Lynx\SpectrumCacheTime
Windows 64-bit: :
HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Canberra Industries,
Inc.\Genie-2000 Environment\Lynx\SpectrumCacheTime
Содержание Lynx DSA
Страница 1: ...Lynx Digital Signal Analyzer 9240227J User s Manual...
Страница 10: ...viii Lynx DSA User s Manual 9240227J Notes...
Страница 232: ...222 Lynx DSA User s Manual 9240227J Notes...