![Burkert FLOWave S 8098 Скачать руководство пользователя страница 90](http://html1.mh-extra.com/html/burkert-0/flowave-s-8098/flowave-s-8098_operating-instructions-manual_2832470090.webp)
90
SAW sensor - Parameter
Type 8098
13.6.3
Activating a user-defined damping of the liquid velocity values
The damping makes it possible to damp the fluctuations of the measured values of the liquid velocity:
On the outputs. The damping of the liquid velocity comes in addition to the damping set for each analog
output (see chpt. 16.3.2 Selecting the damping level of the values transmitted on an analogue output
).*)
By default, the measured liquid velocity values are not damped.
To damp the fluctuations of the measured values, you can:
→
Either choose 1 of the 3 predefined damping levels:
Low
,
Medium
or
High
. See chpt.
→
Or you can set your own damping parameters with the
Special
damping.
With the
Special
damping, you can set 2 parameters:
• a user-defined
Response time
in seconds,
•
the
Jump threshold
, i.e. a user-defined percentage. If 2 consecutive measured values vary for ± the per
-
centage, no damping is applied to the second measured value.
To set your own damping parameters of the measured liquid velocity values, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Parameter
→
Stand. meas. values
→
Liquid velocity
→
Damping
The current settings are displayed.
→
Select
Special
→
Set the value of the
Response time
→
Select if the
Jump threshold
is enabled or disabled
→
If the
Jump threshold
is enabled, set the value.
The new settings are displayed.
→
Finish
The special damping of the liquid velocity values is active.
*
)
Only variant with outputs.
English