137
SAW sensor - Maintenance
Type 8098
15.6
Setting the offset value of the flow zero point
Instead of setting the offset value of the flow zero point, you can calibrate it. See chpt.
offset value of the flow zero point
.
To enter the offset value of the flow zero point, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Maintenance
→
Calibration
→
Stand. meas. values
→
Flow rate
→
Offset
→
Enter value
→
Set the value of the offset. Take into account the direction of the flow as set in chpt.
→
Finish
The offset value of the flow zero point is set.
15.7
Setting the K factor
By default, the value of the K factor is 1.0000.
The K factor can be adjusted, if the measured flow rate values differ from the real values.
Instead of setting the K factor, you can calibrate it by using a teach-in procedure. See chpt.
the K factor by using a teach-in procedure.
To enter the value of the K factor, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Maintenance
→
Calibration
→
Stand. meas. values
→
Flow rate
→
K factor
→
Enter value
→
Set the value of the K factor.
→
Finish
The new K factor value is used.
English