176
Maintenance and troubleshooting
Type 8098
19.8.15
Message "Measure board is in boot starter mode, no firmware
found n°1"
Message
Measure board is in boot starter mode, no firmware found n°1
Symbol displayed
Possible cause
The firmware of the measurement board is lost or is not valid.
What to do?
→
Start the device again.
→
If the error is still there, send the device back to Bürkert.
19.8.16
Message "Data returned by the measurement PCB is invalid
n°1"
Message
Data returned by the measurement PCB is invalid n°1
Symbol displayed
-
Possible cause
The device cannot measure the liquid parameters, for example, because
there are too many bubbles in the liquid or the sensor tube is not completely
filled.
What to do?
→
Make sure there is no problem in the installation.
19.8.17
Message "Communication between transmitter PCB and
measurement PCB has been interrupted n°x"
Message
Communication between transmitter PCB and measurement PCB has been
interrupted n°x
Symbol displayed
Possible cause
There is no communication between the sensor and the transmitter.
What to do?
→
Make sure the cable connecting the sensor to the transmitter is not
broken and correctly plugged in. See chpt.
of the transmitter on the sensor.
19.8.18
Message "The measurement board bootloader operation
failed n° 1"
Message
The measurement board bootloader operation failed n° 1
Symbol displayed
Possible cause
During the firmware update, the sensor software could not be updated.
What to do?
→
Send the device back to Bürkert.