SAW sensor - Parameter
Type 8098
13.4.10
Changing the cut-off value of the flow rate
The default value of the cut-off flow rate is equal to 0.4% of the full scale value. The full scale value depends
on the DN of the measurement tube.
To change the cut-off value of the flow rate, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Parameter
→
Stand. meas. values
→
Volume flow
→
Cut-off
→
Value
→
Set the cut-off value.
→
Finish
The cut-off value of the flow rate is changed.
13.4.11
Disabling the cut-off function
If the cut-off function is enabled, do the following to disable it:
→
Select the device in the navigation area.
→
SAW sensor
→
Parameter
→
Stand. meas. values
→
Volume flow
→
Cut-off
→
Status
→
Select
Disabled
The cut-off function is disabled.