SAW sensor - Maintenance
Type 8098
15.4
Setting the direction of the flow
By default, if the flow direction is opposite the arrow located on the front of the device, the displayed flow
rate values are negative.
If you want that the device displays positive flow rate values, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Maintenance
→
Flow direction
→
Select
Standard
if the arrow located on the front of the device shows the flow direction, or choose
Reverse
if the flow direction is opposite the arrow located on the front of the device.
The flow direction is set and the displayed flow rate values are positive.