184
Maintenance and troubleshooting
Type 8098
19.10.8
Message "Fluid velocity too low"
Message
Fluid velocity too low
Symbol displayed
Possible cause
The value of the liquid velocity is lower than the permitted minimum error
value set in chpt.
13.6.7 Changing the error limits, the warning limits and the
hysteresis of the liquid velocity.
What to do?
As soon as the value of the liquid velocity returns to within the permitted
range, the error is automatically reset.
Message
Fluid velocity too low
Symbol displayed
Possible cause
The value of the liquid velocity is lower than the permitted minimum warning
value set in chpt.
13.6.7 Changing the error limits, the warning limits and the
hysteresis of the liquid velocity.
What to do?
As soon as the value of the liquid velocity returns to within the permitted
range, the warning is automatically reset.
19.10.9
Message "Density factor too high"
Message
Density factor too high
Symbol displayed
Possible cause
The value of the density factor is higher than the permitted maximum error
value set in chpt.
13.8.5 Changing the error limits, the warning limits and the
hysteresis of the density factor.
What to do?
As soon as the value of the density factor returns to within the permitted
range, the error is automatically reset.
Message
Density factor too high
Symbol displayed
Possible cause
The value of the density factor is higher than the permitted maximum
warning value set in chpt.
13.8.5 Changing the error limits, the warning
limits and the hysteresis of the density factor.
What to do?
As soon as the value of the density factor returns to within the permitted
range, the warning is automatically reset.