SAW sensor - Maintenance
Type 8098
15.19
Setting the linearity value of the acoustic
transmission factor
To enter a linearity value for the acoustic transmission factor, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Maintenance
→
Calibration
→
Add. meas. values
→
Acoustic transmission factor
→
Linearity
→
Enter value
→
Set the value of the linearity.
→
Finish
The linearity value of the acoustic transmission factor is set.
15.20
Resetting all the calibration data to the default values
(additional measurement values)
The calibration data that can be reset is:
• the offset value of the density factor
•
the linearity value of the density factor
• the offset value of the acoustic transmision factor
•
the linearity value of the acoustic transmision factor
To reset all the calibration data to the default values, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Maintenance
→
Calibration
→
Add. meas. values
→
Reset to default
→
Finish
.
All the calibration data is reset to the default values.