Maintenance and troubleshooting
Type 8098
19.9.3
Message "Calibration cancelled"
Message
Calibration cancelled
Symbol displayed
Possible cause
The user has interrupted a calibration of the zero flow, before the waiting
time of 30 seconds has elapsed.
What to do?
→
Do a new calibration and observe the given instructions.
19.9.4
Message "Calibration cancelled, the flow rate is less than 5%
of the full scale"
Message
Calibration cancelled, the flow rate is less than 5% of the full scale
Symbol displayed
Possible cause
The calibration has failed because
the flow rate is less than 5% of the full
scale.
What to do?
→
Make sure the flow rate is higher than 5% of the full scale.
→
Do a new calibration.
19.9.5
Message "Resulting K factor is less than 0.8 or higher than
1.2"
Message
Resulting K factor is less than 0.8 or higher than 1.2
Symbol displayed
Possible cause
When calibrating the K factor by using a teach-in procedure depending
on the flow rate or depending on a known volume, you have entered a ref
-
erence value that varies for ±20% from the measured value.
What to do?
→
Do a new calibration.
→
Enter a correct reference value.
19.9.6
Message "Resulting offset is higher than 10 °C, 18 °F"
Message
Resulting offset is higher than 10 °C, 18 °F
Symbol displayed
Possible cause
When calibrating the offset value of the liquid temperature, you have entered
a reference value that varies for ±10 °C (18 °F) from the measured value.
What to do?
→
Do a new calibration.
→
Enter a correct reference value.