Maintenance and troubleshooting
Type 8098
19.11.3
Message "Not totally filled"
Message
Not totally filled
Symbol displayed
Depends on the device status the event is associated to.
Possible cause
The tube is not totally filled. Thus not all the sensors are in contact with the
liquid and measurement is not possible.
The related event must be enabled. See chpt.
special events in the process, on the sensor or on the electronics.
What to do?
→
Make sure the measurement tube is completely filled, for example by
increasing the flow rate.
19.11.4
Message "Liquid out of range"
Message
Liquid out of range
Symbol displayed
Depends on the device status the event is associated to.
Possible cause
The speed of sound in the liquid is out of range.
The related event must be enabled. See chpt.
special events in the process, on the sensor or on the electronics.
What to do?
→
Make sure the liquid in the pipe meets the technical specifications given
in the data sheet for the device and in chpt.
19.11.5
Message "Unstable flow rate"
Message
Unstable flow rate
Symbol displayed
Depends on the device status the event is associated to.
Possible cause
The flow rate is not stable. The standard deviation of the flow rate measure
-
ments is too high.
The related event must be enabled. See chpt.
special events in the process, on the sensor or on the electronics.
What to do?
→
Make sure the operation of the equipment in the process, such as
pumps and process valves, is correct.