13.8.8
Setting the temperature compensation for a liquid other than
water
You can set the 5 constants (a
0
...a
5
) of the equation which compensates the temperature of the liquid to
calculate the density factor:
a
0
+ a
1
T + a
2
T
2
+ a
3
T
3
+ a
4
T
4
+ a
5
T
5
→
To help you define the 5 constants a
0
to a
5
, contact Bürkert.
To activate the temperature compensation for a liquid other than water, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Parameter
→
Add. meas. values
→
Density factor
→
Density factor compensation
→
Density factor compensation
The current settings are displayed.
→
Select
Manual
→
Set the value of the constant a
0
, in the scientific notation. For example, to set the value 0,93724, enter
93.724000E-02 or, to set the value 372,4, enter 3.02.
→
Set the value of the constant a
1
, in the scientific notation.
→
Set the value of the constant a
2
, in the scientific notation.
→
Set the value of the constant a
3
, in the scientific notation.
→
Set the value of the constant a
4
, in the scientific notation.
→
Set the value of the constant a
5
, in the scientific notation.
The new settings are displayed.
→
Finish
The temperature compensation for a liquid other than water is active.