SAW sensor - Parameter
Type 8098
None
damping: all
the fluctuations of the
measured values are
transmitted
Low
damping
Medium
damping
High
damping
Selected damping level is applied
No damping if 2 consecutive measured values vary for ±20 °C
Figure 21:
Operation of the available damping levels
Damping level
Response time
None
0 s
Low
1 s
Medium
10 s
High
30 s
Special
User-defined
Response time
: see chpt. 13.5.3
Table 9:
Response times (10%...90%) of the damping levels for the liquid temperature measurements
To set a predefined damping level of the measured liquid temperature values, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Parameter
→
Stand. meas. values
→
Temperature
→
Damping
The current settings are displayed.
→
Select a damping level between
Low
,
Medium
and
High
The new settings are displayed.
→
Finish
The damping of theliquid temperature values is active and a predefined damping level is selected.