SAW sensor - Parameter
Type 8098
Special event in the process Meaning
Special condition
Low flow cut off
The cut-off value of the flow rate has
been used.
The cut-off function must be
enabled: see chpt.
.
Change of liquid
A different liquid flows in the pipe.
The speed of sound in the liquid
has changed by more than 3 m/s
in 1 second.
Backward flow
The liquid flows in the opposite
direction as the one set in chpt.
Setting the direction of the flow
-
Table 11:
Diagnostics: special events in the process
Special event occurring on the sensor:
Special event on the sensor
Meaning
Special condition
Sound cond. out of range
There are gas bubbles or solid
particles in the liquid.
-
Table 12:
Diagnostics: special events on the sensor
Special event on the electronics:*)
Special event on the
electronics
Meaning
Special condition
Output 1, open loop
There is a connection problem on the
related output.
The related analogue output must
not be disabled. See chpt. 16.4
Output 2, open loop
Output 1, Diag. error
There is a connection problem on the
related output or a high resistance is
detected in the loop.
The related analogue output must
not be disabled. See chpt. 16.4
Output 2, Diag. error
Output 1 overload
An overload has been detected on the
related digital output.
The output has switched.
-
Output 2 overload
Table 13:
Diagnostics: special events on the electronics
*
)
Only variant with outputs.