SAW sensor - Maintenance
Type 8098
15.12
Resetting the offset of the liquid temperature to the
default value
To reset the offset of the liquid temperature to the default value, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Maintenance
→
Calibration
→
Stand. meas. values
→
Temperature
→
Reset to default
→
Finish
The temperature offset is reset to its default value.
15.13
Resetting all the calibration data to its default values
(standard measurement values)
The calibration data that can be reset is:
•
the K factor
• the offset value of the flow zero point
• the offset value of the liquid temperature
To reset all the calibration data to the default values, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Maintenance
→
Calibration
→
Stand. meas. values
→
Reset to default
→
Finish
All the calibration data is reset to the default values.