56
General settings - Parameter
Type 8098
10.7.1
Reading out the 2 error limit values
To read out the limits the supply voltage of the device should be in, do the following:
→
Select the device in the navigation area.
→
General settings
→
Parameter
→
Alarm limits
→
Supply voltage
or
Device temperature
Error high
or
Error low
can be read.
10.7.2
Changing the 2 warning limit values
To change the warning limits of the supply voltage or of the device temperature, do the following:
→
Select the device in the navigation area.
→
General settings
→
Parameter
→
Alarm limits
→
Supply voltage
or
Device temperature
→
Warning high
or
Warning low
→
Set the warning limit.
The warning limits are changed.
10.7.3
Reading out the hysteresis value
To read out the hysteresis value, do the following:
→
Select the device in the navigation area.
→
General settings
→
Parameter
→
Alarm limits
→
Supply voltage
or
Device temperature
Hysteresis
can be read.
English