![Burkert FLOWave S 8098 Скачать руководство пользователя страница 88](http://html1.mh-extra.com/html/burkert-0/flowave-s-8098/flowave-s-8098_operating-instructions-manual_2832470088.webp)
88
SAW sensor - Parameter
Type 8098
13.6
Setting the parameters of the liquid velocity
13.6.1
Giving a user defined name to the measured liquid velocity
The name is used to identify the process value in the user defined views and in all the menus where the
process value is displayed (for example in the
Outputs
menu).
By default, the name associated to the measured liquid velocity is
Liquid velocity
.
To add a user defined name to the default name, do the following:
→
Select the device in the navigation area.
→
SAW sensor
→
Parameter
→
Stand. meas. values
→
Liquid velocity
→
Value name
→
Enter the name. The name can have up to 19 characters.
→
Apply
The name is changed.
13.6.2
Activating the damping of the liquid velocity values and
selecting a predefined damping level
The damping of the liquid velocity comes in addition to the damping set for the volume flow. The damping
makes it possible to damp the fluctuations of the measured values of the liquid velocity:
On the outputs. The damping of the liquid velocity comes in addition to the damping set for each analog
output (see chpt. 16.3.2 Selecting the damping level of the values transmitted on an analogue output
).*)
The damping is not applied to the new measured value, if the 2 following conditions are met:
•
a
Low
,
Medium
or
High
damping level is active
• and the variation between 2 values that are measured one after the other is higher than 30 % (for example
when charging the pipe or stopping the flow)
By default, the liquid velocity values are not damped.
→
As an alternative to the 3 predefined damping levels
Low
,
Medium
or
High
, you can set your own
damping parameters. See 13.6.3.
*
)
Only variant with outputs.
English