![Breas Medical Vivo 45 Скачать руководство пользователя страница 87](http://html1.mh-extra.com/html/breas-medical/vivo-45/vivo-45_user-manual_2806367087.webp)
87
How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
Using the Vivo 45 with the FiO
2
Sensor
The FiO
2
sensor can be used to monitor and store FiO
2
measurements.
The FiO
2
sensor measures the fraction of inspired oxygen in the air chan-
nel of the Vivo 45. The FiO
2
measurements will be stored in the data
memory which can be downloaded to a PC and viewed in Breas software.
U
SAGE
T
IME
Operating temperature
10 to 40°C (50 to 104°F)
Operating pressure
700 to 1250 mbar
Expected operating life
<6 years (in ambient air)
1 year (in 100% O
2
)
Shelf life
6 months
Note that the operating conditions for the FiO
2
sensor is different from
the Vivo 45 system conditions. If the sensor is used outside its operating
conditions the FiO
2
measurements might deviate.
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...