186 Technical Specifications
Vivo 45 User Manual
Doc.006149 En Y-2
Filtering/Smoothing Techniques
F
UNCTION
T
ECHNIQUE
D
ESCRIPTION
Pressure
Low pass average time constant 16 ms
Inspiration trigger
Differential mass flow resolution 4 ms
Expiration trigger
Flow low pass filtering with level sensing
SpO
2
No data post-processing done by the ven-
tilator
Effort Belt
Low pass filter: 5Hz
High pass filter: 0.1Hz