110 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
High PEEP Alarm
I
TEM
D
ESCRIPTION
Definition
A High PEEP alarm will be given when the meas-
ured PEEP is 30% above the set value for more
than 15 ±0.5 seconds.
Priority
Medium
Possible cause
• Too short expiratory time.
• Changes in airway resistance and or compli-
ance.
• Blocked leakage port.
Setting range
On, Off
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication