69
How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
Max Pressure
I
TEM
D
ESCRIPTION
Definition
The Max Pressure setting is only used when
Target Volume is activated. Max Pressure
defines the upper pressure limit up to where
the Vivo 45 can increase the pressure to
reach the set Target Volume. If Target Vol-
ume is not reached at Max Pressure, the
Vivo 45 will continue to ventilate at this Max
Pressure setting.
Setting min
Min Pressure
Setting max
40 cmH
2
O
Setting resolution
0.5 below 10 cmH
2
O,
1.0 above 10 cmH
2
O
Tolerance
±0.5 cmH
2
O or ±5 %, whichever is the
greatest.