Inspiratory Time (Insp. Time)
Backup Inspiratory Time (Backup Insp. Time)
I
TEM
D
ESCRIPTION
Definition
The Inspiratory Time setting defines the
length of each inspiration from start of inspi-
ration to cycling off to expiration.
The combination of the Inspiratory Time
and Breath Rate settings is limited by the I:E
ratio 2:1.
Setting min
0.3 s
Setting max
5 s (Adult), 2 s (Paediatric)
Setting resolution
0.1 s
Display
I
TEM
D
ESCRIPTION
Definition
The Backup Inspiratory Time setting defines
the length of each inspiration delivered dur-
ing ventilator-triggered backup ventilation,
initiated by the set Backup Rate.
The combination of the Backup Inspiratory
Time and Backup Rate setting is limited by
the I:E ratio 2:1.
Setting min
0.3 s
Setting max
5 s (Adult), 2 s (Paediatric)
Setting resolution
0.1 s
Display