126 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
High EtCO
2
Alarm
I
TEM
D
ESCRIPTION
Definition
A High EtCO
2
alarm will be given when the
measured EtCO
2
exceeds the alarm limit for
30 seconds.
Priority
High
Possible cause
• Alarm limit is set too low.
• Breath Rate too low.
• Delivered Tidal Volume too low.
• Excessive dead space between patient and
leakage port.
• Leak port occluded.
Setting min
1 mmHg
Setting max
74 mmHg, Off
Setting resolution
1 mmHg
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication