62 How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
PEEP
Breath Rate
I
TEM
D
ESCRIPTION
Definition
The PEEP setting is used to define the air-
way pressure at the end of the expiratory
phase.
Setting min
2 cmH
2
O
Setting max
20 cmH
2
O, Inspiratory Pressure -2 cmH
2
O
or Min Pressure -2 cmH
2
O.
Setting resolution
0.5 below 10 cmH
2
O,
1.0 above 10 cmH
2
O
Display
I
TEM
D
ESCRIPTION
Definition
The Breath Rate setting defines the mini-
mum number of breaths the Vivo 45 will
deliver as long as no inspiratory trigger
effort from the patient is detected. The
cycles will be ventilator-initiated breaths.
The combination of the Breath Rate and
Inspiratory Time setting is limited by the I:E
ratio range 1:9.9 to 2:1.
Setting min
4 bpm (Adult), 6 bpm (Paediatric)
Setting max
40 bpm (Adult), 60 bpm (Paediatric)
Setting resolution
1 bpm
Display