138 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Lost Mains Power Alarm
I
TEM
D
ESCRIPTION
Definition
A Mains Power Lost alarm will be given when
the Vivo 45 switched from Mains to another
power source due to Mains power is lost.
Priority
Medium
Ventilator action
The Vivo 45 will continue treatment with the
same settings. An information message will
be shown on the screen.
Indication