136 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Low Last Power Source Alarm
I
TEM
D
ESCRIPTION
Definition
A Low Last Power Source alarm will be given
when the last battery source (internal battery
or click-in battery) has 15 to 20 minutes of
operating time left with current settings.
Priority
Medium
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication