118 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Apnea Alarm
I
TEM
D
ESCRIPTION
Definition
An Apnea alarm will be given when no
patient-triggered breath is detected for the
set period of time.
This alarm is only used in assisted breath
modes.
Priority
High
Possible cause
• Inspiratory Trigger is set too high.
• Patient stopped breathing.
• Patient decreases spontaneous breathing.
• Circuit disconnection.
Setting min
Off, 5 s
Setting max
60 s
Setting resolution
5 s below 15 s,
15 s above 15 s.
Ventilator action
The Vivo 45 will continue treatment with
the same settings.
Indication