122 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
High FiO
2
Alarm
I
TEM
D
ESCRIPTION
Definition
A High FiO
2
alarm will be given when the meas-
ured FiO
2
exceeds the alarm limit for
30 ±0.5 seconds.
Priority
Medium
Possible cause
• Increased oxygen inflow.
• Decreased minute ventilation.
Setting min
21%
Setting max
100%, Off
Setting resolution
1%
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication