109
Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Low Pressure Alarm
I
TEM
D
ESCRIPTION
Definition
A Low Pressure alarm will be given when the
Vivo 45 pressure fails to reach the low pressure
alarm limit for 15 ±0.5 seconds.
Priority
High
Possible cause
• Disconnection of patient circuit.
• Mismatch between Inspiratory pressure/
CPAP and alarm setting.
• Leakage from the mask or other components
of the patient circuit.
Setting min
1 cmH
2
O
Setting max
40 cmH
2
O
Setting resolution
0.5 below 10 cmH
2
O, 1.0 above 10 cmH
2
O
The Low Pressure alarm setting is also visible
with a red line in the pressure bar graph.
Indication