71
How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
Ramp Time
Start Pressure
Always Start Ramp
I
TEM
D
ESCRIPTION
Definition
The Ramp Time setting defines the time
during which the pressure will increase from
the set Start Pressure to the set Inspiratory
Pressure.
Ramp Time is only available in Clinical
Mode.
Setting min
Off
Setting max
60 min
Setting resolution
10 min
I
TEM
D
ESCRIPTION
Definition
The Ramp Start Pressure defines the pres-
sure level delivered by the ventilator at the
beginning of the Ramp Time.
Setting min
2 cmH
2
O
Setting max
Insp. pressure – 2 cmH
2
O
I
TEM
D
ESCRIPTION
Definition
The Always Start Ramp determines whether
the ramp is started automatically at the
beginning of the treatment.
Setting min
Off
Setting max
On