132 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
PtcCO
2
Alarm
I
TEM
D
ESCRIPTION
Definition
A PtcCO
2
alarm will be given when PtcCO
2
is
outside alarm limits.
Check the PtcCO
2
monitor.
Priority
High
Possible cause
• External PtcCO
2
monitor is outside its alarm
limits.
• Breath Rate needs adjustment
• Delivered Tidal Volume needs adjsutment.
• Excessive dead space between patient and
leakage port.
• Leak port occluded.
• Ventilator disconnection.
• Excessive leakage in the Patient circuit/Inter-
face.
• Partial obstruction of the airways.
• Self triggering of the ventilator.
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Indication