41
Preparing the Vivo 45 for Use
Vivo 45 User Manual
Doc.006149 En Y-2
4.7 Performing the Pre-use Test
The pre-use test is used for detecting the characteristics of the patient cir-
cuit that is connected to the Vivo 45. The resistance and compliance of
the patient circuit are measured and calculated. This will be used to com-
pensate for pressure drop in the patient circuit and the compliance of the
patient circuit.
The patient shall not be connected during the pre-use test.
The compliance is used for correcting the volume of air delivered to the
patient by subtracting the air retained in the patient circuit during pressur-
ization.
Starting the Pre-use Test Manually
1
On the
Others
menu, Select
Pre-use Test
and then
Start Pre-use
Test
.
2
Follow the instructions on the display.
Activating the Pre-use Test Prompt
This activates a prompt to perform a pre-use test every time the Vivo 45
is powered on.
1
On the
Others
menu, Select
Pre-use Test.
2
Set P
re-use Test
to
On
.
Always perform a new pre-use test if the patient circuit configuration has
been changed.
If the pre-use test has not been performed, the Vivo 45 will operate with
default patient circuit compensation.
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...