168 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Low Pressure and Disconnection Alarms
1
Start treatment and disconnect the patient circuit.
2
Wait 15 seconds.
3
The Low Pressure Alarm and/or the Disconnection Alarm will be
given.
4
Stop treatment. Test completed.
Low Vte Alarm
1
Start treatment and block the patient circuit completely to simulate
an occlusion.
2
Wait 15 seconds.
3
The Low Vte Alarm will be given.
4
Stop treatment. Test completed.
Obstruction Alarm
1
Start treatment; block the patient circuit completely to simulate an
obstruction.
2
Wait approximately 10 seconds.
3
The Obstruction Alarm will be given.
4
Stop treatment. Test completed.