68 How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
Target Volume
I
TEM
D
ESCRIPTION
Definition
The Target Volume setting defines the tidal
volume that the Vivo 45 will aim for while
ventilating the patient in a pressure mode. To
aim for the preset volume, the Vivo 45 will
adapt the Inspiratory Pressure between two
adjustable pressure limits: Min Pressure and
Max Pressure.
When Target Volume is active, the mode
field on the Vivo 45 display will indicate
“(TgV)”.
Setting min
Off, 300 ml (Adult), 100 ml (Paediatric)
Setting max
2000 ml (Adult), 500 ml (Paediatric)
Setting resolution
10 below 500 ml,
50 above 500 ml
Display