65
How to Use the Vivo 45
Vivo 45 User Manual
Doc.006149 En Y-2
Inspiratory Trigger (Insp. Trigger)
I
TEM
D
ESCRIPTION
Definition
The inspiratory trigger setting defines the
patient’s effort required to initiate a ventila-
tor assisted breath. When the patient starts a
breath, an increasing flow is created in the
patient circuit. If the patient’s effort reaches
the set inspiratory trigger level an inspiration
is initiated. If the patient cannot trigger a
breath, the ventilator will deliver breaths
according to the set Backup Rate or Breath
Rate.
The Assisted breath modes in PCV are
turned off if the inspiratory trigger is set to
Off.
Setting min
1
Setting max
9 (PSV),
9, Off (PCV)
Setting resolution
1
(Setting 1 is the most sensitive and 9 is the
least sensitive)
Display
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...