![Breas Medical Vivo 45 Скачать руководство пользователя страница 106](http://html1.mh-extra.com/html/breas-medical/vivo-45/vivo-45_user-manual_2806367106.webp)
106 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
• Alarm text in display: displays
the name of the active alarm
condition.
If several alarm conditions have
been reached, the alarm descrip-
tions are rolling in the display,
displaying the alarms with high-
est priority first.
A “>>” symbol is indicating
that more than 1 alarm is set.
• Audible signals: indicate the priority of the active alarm condition.
– High priority: 3 signals followed by 2 more. The signal sequence is
repeated with a 0.5 second pause and thereafter a 3 second pause.
– Medium priority: 3 signals, with a lower frequency than the high prior-
ity alarm. The signal sequence repeats after a 6 second pause.
– Information: 1 signal with a low frequency. The signal is repeated after
a 5 second pause and stopped after 5 sequences.
– Function failure. Same signal as the high priority alarm or a constant
signal, depending on the kind of function failure.
• The power failure alarm sounds in the case of power failure.
• If a battery that is the last power source falls below the warning limit,
the Low Last Power Source alarm is set.
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...