167
Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
14.20
Start treatment and wait 30 s.
14.21
Low pulse rate alarm should be given.
14.22
Stop Treatment.
14.23
Set the low pulse rate alarm to off.
15
Alarm test completed.
Additional Optional Alarm Tests
In this chapter, methods for additional alarm tests are described. These
tests are optional and not needed to ensure safe use of the Vivo 45.
High PEEP Alarm
1
Connect the Vivo 45 patient circuit to a test lung and a CPAP device.
2
Set the CPAP device treatment pressure to 10 cmH
2
O.
3
Adjust the Vivo 45 settings as follows:
4
Start treatment on both the Vivo 45 and the CPAP device.
5
Wait approximately 15 seconds before the High PEEP alarm shall be
given.
6
Stop treatment. Test completed.
S
ETTING
V
ALUE
Ventilation Mode
Pressure
Breath Mode
Assist/Control
Patient Mode
Adult
Insp. Pressure
15 cmH
2
O
PEEP 5 cmH
2
O
Breath Rate
12 bpm
Insp. Time
1.5 s
Rise Time
5
Insp. Trigger
Off
Target Volume
Off
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...