![Breas Medical Vivo 45 Скачать руководство пользователя страница 164](http://html1.mh-extra.com/html/breas-medical/vivo-45/vivo-45_user-manual_2806367164.webp)
164 Alarms
Vivo 45 User Manual
Doc.006149 En Y-2
Humidifier/Bypass Loose Alarm
I
TEM
D
ESCRIPTION
Definition
The Humidifier/Bypass Loose alarm shall be
given when the air bypass/humidifier is not
inserted correctly.
Priority
Medium
Ventilator action
The Vivo 45 will continue treatment with the
same settings.
Reset action
Reinsert the air bypass unit/humidifier and
make sure the latch closes.
Indication
Содержание Vivo 45
Страница 10: ...10 Introduction Vivo 45 User Manual Doc 006149 En Y 2...
Страница 131: ...131 Alarms Vivo 45 User Manual Doc 006149 En Y 2...
Страница 176: ...176 Cleaning and Maintenance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 196: ...196 Technical Specifications Vivo 45 User Manual Doc 006149 En Y 2...
Страница 206: ...206 Accessories and Parts Vivo 45 User Manual Doc 006149 En Y 2...
Страница 208: ...208 Patient Settings Vivo 45 User Manual Doc 006149 En Y 2...
Страница 210: ...210 FAA Compliance Vivo 45 User Manual Doc 006149 En Y 2...
Страница 226: ...226 Index Vivo 45 User Manual Doc 006149 En Y 2...