Alignment
Straightness of the Sample or Reagent Probe (from side)
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-54
September 2016
Straightness of the Sample or Reagent Probe (from side)
Follow the steps below to verify the "straightness" (trueness) of the Sample Probe
viewing it from the side of the probe.
Step
Action
1
Disable all Electro-Mechanical Motion. Refer to Home All and Disable
All in this chapter.
2
If aligning the Sample Probe,
follow the steps below...
If aligning the Reagent Probe,
follow the steps below...
Remove the sample carousel
cover by:
• removing the two hex screws
(1) on top of the carousel
cover.
• removing the two Phillips
screws (2) on the right side just
above the fan.
Open the Reagent Carousel lid.
Place a piece of tape along the
top left edge of the sample
carousel tub. (Refer to Figure
10.25.)
Place a piece of tape along the
top right edge of the reagent
carousel tub. (Refer to Figure
10.26.)
(1 of 6)