Calibration Results
Introduction
Reagents/Calibration
IMMAGE 800 Operations Manual A11403
Page 6-34
September 2016
Calibration Results
Introduction
The following information may be printed on the Calibration Report for a successful
calibration result: (Refer to APPENDIX B,
for an example report.)
• Chemistry ID
• Chemistry Lot
• Calibrator ID
• Reaction Buffer Lot
• Sample Diluent Lot
• Date and Time of Calibration
• Units
• Instrument Response
• Target Response
• Cal Value
• Dilution
• Rack/Position of the calibrator
• Reagent Position
• STAT/Rerun flags
• Instrument Errors/Codes
• Calibrated to Value
Calibration Time Limit
No time limit is imposed by the system for expiration of a calibration result.
Recalibration of affected reagents may be necessary when reaction buffer or diluent
lots are changed. (Refer to the IMMAGE Immunochemistry Systems
Chemistry
Information Manual
.)
After Calibration
Rack positions used for calibration are automatically cleared to a status of
Available
after the calibration is run whether the calibration passed or failed.
Failed Calibration
When calibration fails:
• Samples programmed for a reagent that fails calibration will be displayed and
printed on the report as
Pending
.
• Refer to CHAPTER 10,
Utilities