Errors
Error Table
Utilities
IMMAGE 800 Operations Manual A11403
Page 10-30
September 2016
804
Level Sense
Failure at Diluent
Bottle
WARNING:
1. Check and remove bubbles from
diluent.
2. Rerun sample.
805
Level Sense
Failure at
Dilution Well
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
806
Level Sense
Failure at Sample
Position
WARNING:
1. Check and remove bubbles from
sample.
2. Verify sufficient sample volume is
present.
3. Rerun sample.
811
Level Sense
False Trigger at
Buffer Bottle
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
812
Level Sense
False Trigger on
Reagent A at
Position
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
813
Level Sense
False Trigger on
Reagent B at
Position
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
814
Level Sense
False Trigger at
Diluent Bottle
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
815
Level Sense
False Trigger at
Dilution Well
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
816
Level Sense
False Trigger at
Sample Rack
WARNING:
CALL BECKMAN COULTER
CLINICAL SUPPORT
900
Software
Initialization
Error
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
901
Error
Downloading
DAS
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(9 of 21)