Sample and Reagent Crane Probe
Replacing Sample and Reagent Crane Probe
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-75
10
7
At the crane assembly, remove the small set screw (1) located near the top,
right side of the probe RF (radio frequency) shield housing (2) (a brass
cylinder). (Refer to Figure 10.41.)
Figure 10.41
8
Pull the probe up and out from the crane assembly. (Refer to Figure
10.42.)
Figure 10.42
Step
Action
, continued
(4 of 6)
2
1
A014086P.EPS
A012191P.EPS