Selecting Chemistry Tests by Panel
After Panel Selection
IMMAGE 800 Operations Manual A11403
Sample Programming
September 2016
Page 7-17
7
Figure 7.3
Panels Dialog Box
After Panel Selection
After a panel is selected:
• the panel number will remain displayed.
• the chemistries in that panel will be selected in the Chemistry Menu.
• the sample type defined for the panel will be displayed.
• any off-line or non-standard dilution defined for the panel will be applied.
Canceling a Panel
A selected panel can be canceled by deselecting it in the Panels list. The highlight will
be removed.
A panel cannot be canceled after the sample program is saved and/or the run is started.
Additional Information
Refer to CHAPTER 5,
System Setup
, Panel Setup for more information on defining
panels.
E010253S.EPS
NOTICE
If multiple panels are selected, their defined sample types non-standard dilutions,
antigen excess and any off-line dilutions must be the same.