Defining a UDR Chemistry
Description of Definition Fields
System Setup
IMMAGE 800 Operations Manual A11403
Page 5-76
September 2016
Figure 5.34
The Reagent Compartment
The following table describes the calibration fields of the Define/Edit User-Defined
Chemistry screen, Page 3. (Refer to Figure 5.35.)
A011410P.EPS
2
3
4
5
1
6
7
7
B
A
1. Reagent Compartment Cover
2. Reagent Carousel
3. Reaction Buffer Bottle
4. Reagent Cartridges (Compartments A and B)
5. Reagent Bar Code Reader
6. Fans
7. Temperature Sensor
Table 5.9
Calibration Definition Fields, Page 3
Field
Entries Allowed
Function
Levels
4 to 9
Identifies the number of
calibrators used in test.
Replicates
1 to 9
Allows a number of tests to be
repeated for each calibration
level.
Update Level
1 to 9
Single-point calibration setpoint
update level.
Replicates
1 to 9
Set point update replicate.
Cal Level Setpoints
Up to 6 digits with a
decimal point, or seven
digits.
Identifies concentration value
for each calibration level in
ascending order where Level 1
is the lowest concentration
value.