Errors
Error Table
IMMAGE 800 Operations Manual A11403
Utilities
September 2016
Page 10-25
10
E36
Sample LVL
Level Sense Failure on
Sample
1. Check and remove bubbles from
sample.
2. Verify sufficient sample volume is
present.
3. Rerun sample.
E37
RGT A N/A
Reagent Compartment A
Unavailable
Calibrate reagent.
E38
RGT B N/A
Reagent Compartment B
Unavailable
Calibrate reagent.
E39
Buff N/A
Buffer Unavailable
Load a new buffer bottle, and verify
that buffer position is defined
correctly.
E40
Dil N/A
Diluent Unavailable
Load a new diluent bottle, and verify
that diluent position is defined
correctly.
E41
Hrdwr N/A
Hardware Unavailable
Home system.
E42
Excess Scat
Excessive Scatter
CALL BECKMAN COULTER
CLINICAL SUPPORT
E43
Blank Term
Assoc. Blank Terminated CALL BECKMAN COULTER
CLINICAL SUPPORT
E44
S/W Error
Bad Test Object Address CALL BECKMAN COULTER
CLINICAL SUPPORT
E45
Uncal RGT
Uncalibrated Reagent
Recalibrate reagent.
E46
DilWell Err
Dilution Segments not
Defined or Cleared
1. Clear dilution segments.
2. Rerun sample.
E47
S/W Error
Scheduling Failed
CALL BECKMAN COULTER
CLINICAL SUPPORT
E48
No Param
No Valid Reagent
Parameters A-G
1. Reload reagent card.
2. Reload reagent.
E49
Buff N/A
Insufficient Buffer
Load a new buffer bottle, and verify
that buffer position is defined
correctly.
50
Process
Exception
FATAL:
CALL BECKMAN COULTER
CLINICAL SUPPORT
Table 10.6
Displayed or Printed Error Messages
, continued
Error #
Abbreviation/
Message
Explanation
Action
(4 of 21)